Nitrous oxide cartridges for CryoPen®
CryoPen B+: 23.5 g cartridges
CryoPen X+: 8 g and 16 g cartridges
CryoPen XP: 8 g and 16 g cartridges
CryoPen O: 8 g and 16 g cartridges
Using the instrument:
Step 1: Loading a new cartridge
The cartridges are packaged in peelable thermoformed blister packs with an assembled filter/piercer. The filters are used to protect the applicator, and the piercer is used to pierce the cartridge.
CryoPen X+ and CryoPen XP can be used with 16 g N₂O cartridges as well as 8 g cartridges, thanks to the 8/16 g converter. Check that the cavity contains no remaining elements.
Open the peelable packaging. Remove the filter/piercer with the tweezers. Hold the CryoPen with the end pointing downwards. Place the filter/piercer in the instrument, ensuring it lies flat at the bottom of the loading cavity.
The piercer trocar must point upwards! Take the cartridge and insert it into the cavity, narrow end downwards, above the filter/piercer.
SEE IMAGE 1
Step 2: Piercing the unused cartridge
Carefully screw the rear end of the CryoPen onto the front end.
Continue until you feel that contact has been established between the filter/piercer block and the cartridge. Then position your hands so that you can make one final, uninterrupted rotation. Then turn quickly until you reach the stop.
You may hear a brief hiss.
The briefer, the better!
Removing the used cartridge and emptying the cavity
Before loading a new cartridge, first release all residual gas from the cartridge by unscrewing the applicator and then pressing the trigger. Then unscrew the rear end. Discard the used cartridge and filter/piercer.
For CryoPen X+ and CryoPen XP instruments, be careful not to discard the 16/8 g converter.
Make sure that the used filter/piercer does not remain in the cavity. The cavity must no longer contain any elements.
Step 3: Using your CryoPen
Remove the applicator cap. Gently press the trigger to start the N₂O flow. Release the trigger to stop the flow of cryogenic liquid.
1 or 2 seconds after releasing the trigger, the gas remaining in the applicator is expelled.
One cartridge can be used for several treatments and patients.
Important notes:
- Make sure to use only cartridges and filters/piercers supplied by H&O Equipments nv/sa. Any other cartridge or filter/piercer will damage the instrument.
- Incorrect positioning of the filter/piercer device in the loading cavity will cause irreversible damage to the instrument.
- Residual foreign bodies present in the loading cavity when inserting a new filter/piercer may cause the instrument to fail.
- The loaded instrument is under high pressure. Any modification made to the structure of the instrument may pose a potential risk.
- Never obstruct the gas expulsion opening with your hand at the end of the rear part when removing the cartridge!
- Cartridges must be kept away from heat (max 50°C/122°F). Leave the applicator cap on when the applicator is not in use.
- CryoPen instruments are intended exclusively for professional use.
- N₂O cartridges contain both a liquid phase and an equilibrium gas phase. The proportion of each phase depends on the ambient temperature and affects the output flow. Optimal performance is achieved when the cartridge temperature is between 19°C and 22°C. Temperatures below or above this range will result in different pressure and density values for the cryogenic substance. Above 25°C, the amount of liquefied N₂O in the CryoPen jet will decrease.
- Empty cartridges must be discarded with metal waste.
The user manual is available in the "attachments" tab above